Model Number 16-02-85 |
Device Problem
Electrical /Electronic Property Problem (1198)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/25/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
There was no patient involvement.Livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in united states.The field service representative in charge replaced the circulator motor and distribution board device was test extensively.Performed functional verification tests successful.Unit was released to customer.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova deutschland received a report that a heater-cooler system 3t started smelling like it was on fire during cleaning it.No error codes was displayed.There was no patient involvement.
|
|
Manufacturer Narrative
|
H10: complaints database analysis revealed that no similar event on this device occurred since its installation in 2015.A short circuit most likely caused the reported event; indeed, an electrical malfunction of the replaced parts (circulator motor and distribution board) cannot be excluded.
|
|
Event Description
|
See initial report.
|
|