The issue reported was that while moving the table up in the vertical direction for scan preparation, the table descended rapidly approximately 30cm and when it reached its lowest point there was a loud noise.A philips field service engineer (fse) went onsite to evaluate the system and found the table in its lowest (down) position.The system was inspected and the fse confirmed that the reported problem was due to a defective vertical spindle of the ball screw assembly which caused the table to descend.There was no harm to patient or operator and no medical intervention was necessary as a result.The entire ball screw assembly was replaced and no further deviations were found.The system was tested successfully and returned to the customer for use.In a similar incident, a philips fse collected the vertical ball screw and bearing parts and returned them to the suzhou (sz) factory for a failure analysis.Analysis confirmed the vertical ball screw was fractured and a further material analysis was performed by a 3rd party lab (suzhou metal service co.Ltd).According to the analysis, the ball screw failure was caused by a fatigue fracture.In addition to the analysis, ball screw inventory was checked and the parts met drawing specifications.Therefore, a material and supplier quality issue was excluded for failure root cause.Couch manufacturing process for ball screw alignment and installation were also checked through tolerance stack up and production couch alignment accuracy measurement.The calculation determined the production alignment meets specification.Sampling checked the production couch measurement accuracy, and all were within specifications.Therefore, couch manufacturing was excluded for failure root cause.After reviewing all event details, philips engineering confirmed there were no new failure modes.During site inspection, the philips fse found the upgrade kit was not installed on the couch in the previous ball screw replacement and the alignment of the ball screw replacement was not performed per the service manual; the alignment tool was designed to avoid misalignment during ball screw replacement.Additionally, the philips ct system is intended to be used and operated only in accordance with the safety procedures and operating instructions given in the instructions for use for the purpose for which it was designed.Operators of the philips system must have received adequate training on its safe and effective use before attempting to operate the equipment described in the instructions for use.The philips systems should not be used if any of the following conditions exist or are thought to exist.The preventative maintenance program is not up-to-date.If any part of the equipment or system is known (or suspected to be) operating improperly.During all movements of the gantry and the patient table (automatic and manual), keep the patient under continuous observation.Make sure that the patient is strapped securely to avoid dangling of the hands.Ensure that the patient is placed securely on the patient table and is not in danger of falling.In conclusion, the root cause of the ball screw failure was determined to be fatigue due to misalignment from no use or partial use of the ball screw service install kit.A review of the risk management file indicates the issue reported by the customer is of low potential severity, which would not reasonably cause or contribute to death or serious injury if the problem were to reoccur.Therefore, based on the investigation¿s conclusion, this issue has been determined not to be a reportable event.
|