Model Number 350P |
Device Problem
No Audible Prompt/Feedback (2282)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/25/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Heartsine has requested return of the device for investigation.Upon completion, the conclusions will be submitted in a follow-up report.
|
|
Event Description
|
The customer reported to heartsine that their device does not emit audio.Absence of voice prompts may lead to an inability understanding how to use the device correctly.There was no report of patient use associated to the reported event.
|
|
Manufacturer Narrative
|
Heartsine's investigation of the device confirmed the reported fault as upon receipt the device was intermittently issuing audio prompts.This fault was attributed to the speaker.The device was scrapped by heartsine and the customer was provided with a replacement device.
|
|
Event Description
|
The customer reported to heartsine that their device does not emit audio.Absence of voice prompts may lead to an inability understanding how to use the device correctly.There was no report of patient use associated to the reported event.
|
|