RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number V60 V60PLUS VENTILATOR
Device Problem Volume Accuracy Problem (1675)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/03/2023
Event Type  malfunction  
Manufacturer Narrative
E1: reporting institution name: (b)(6).Reporting address state: (b)(6).Reporting address postal: (b)(6).Reporting institution phone #: (b)(6).Reporter phone #: (b)(6).
 
Event Description
Philips received a complaint by the customer on the v60, indicating that the tidal volume could not be displayed due to a patient circuit leakage.The device was in use on a patient at the time the reported issue was discovered; however, there was no harm to the patient or user.The authorized service provider (asp) engineer confirmed that after the device was turned on, the tidal volume could not be displayed--after reconnecting the power cord, the device turned on, but the fault remained.The asp engineer found that the patient circuit leaked, replaced the patient circuit, and verified that the issue was resolved.The device passed required performance verification tests per philips standard and was returned to service.The investigation concludes that no further action is required at this time.Although requested, the defective parts were not received at this time.A supplemental report will be submitted if the part is received and evaluated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key17000120
MDR Text Key315920968
Report Number2518422-2023-12101
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838025776
UDI-Public(01)00884838025776
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/24/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60 V60PLUS VENTILATOR
Device Catalogue Number1076716
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/03/2023
Date Device Manufactured03/25/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-