The test strips were requested for investigation and have not been received at this time.If the product is returned in the future, a follow-up report will be submitted. the reporter's meter was provided for investigation where it was tested using retention strips and retention controls.Testing results (qc range = 2.3 - 3.5 inr): qc 1: 2.9 inr.Qc 2: 2.8 inr.Qc 3: 2.9 inr.The obtained qc values were in the allowed range of the used combination strip lot - qc lot.All measurements were without error messages.On a regular basis, coaguchek strips of lots currently valid in the market are tested as part of routine retention testing and the results have passed the internal inspection. per product labeling: "coaguchek method uses human recombinant thromboplastin.Therefore, the comparability to tests using other human recombinant thromboplastins is best, whereas higher deviations can occur with other thromboplastin types.However, those higher differences between thromboplastins of different (rabbit, bovine) origin are not an issue specific for coaguchek assays.Similar differences can be observed when a human recombinant thromboplastin-based laboratory method is compared with several other (rabbit, bovine) laboratory methods the investigation did not identify a product problem.The cause of the event could not be determined. section e3 - occupation: patient/consumer h3 other text : na.
|