Model Number 71953-01
Device Problem Product Quality Problem (1506)
Patient Problem Burn(s) (1757)
Event Date 04/22/2023
Event Type  Injury  
Event Description
Customer reported her adc device reader.Customer reported receiving second and third degree burns on both hands after the reader exploded.Customer was hospitalized and received treatment of felodipine for the second and third degree burns and an insulin injection from a healthcare professional.It is unknown at this time if the charging cable and adapter used was the cable and adapter supplied by adc for use with the libre device.There was no report of death, serious injury, or mistreatment associated with this event.Issues involving thermal events occurring with adc freestyle libre and libre 2 readers is associated with report of corrections and removal number 2954323-02/09/23-001-c, submitted to the fda on 09-feb-23.Adc field action fa1005-2023 was issued to the us 09feb23.
 
Manufacturer Narrative
The customer's product has been requested for investigation.A follow-up report will be submitted once additional information is obtained.The device mfg date is unknown.The date entered in section h4 is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
 
Manufacturer Narrative
Repeated attempts by adc to retrieve the product were unsuccessful and/or the customer discarded the product.The most probable root causes associated with this failure mode are disconnected, faulty or damaged components or misuse.However, mitigations are in place to reduce and prevent such issues.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.At this time product has not yet been returned and a valid serial number has not been provided.An extended investigation has been performed for the reported complaint and it has been determined that there was no indication that the product did not meet specification.A tripped trend review was conducted for the reported complaint and libre reader, no trends were identified that would indicate any product related issues.If the product is returned, a physical investigation will be performed and a follow up report submitted.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
Customer reported her adc device reader.Customer reported receiving second and third degree burns on both hands after the reader exploded.Customer was hospitalized and received treatment of felodipine for the second and third degree burns and an insulin injection from a healthcare professional.It is unknown at this time if the charging cable and adapter used was the cable and adapter supplied by adc for use with the libre device.There was no report of death, serious injury, or mistreatment associated with this event.Issues involving thermal events occurring with adc freestyle libre and libre 2 readers is associated with report of corrections and removal number 2954323-02/09/23-001-c, submitted to the fda on 09-feb-23.Adc field action fa1005-2023 was issued to the us 09feb23.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 2
Type of Device
FLASH GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda, CA 94502-7001
5107495297
MDR Report Key17007657
MDR Text Key315991146
Report Number2954323-2023-21788
Device Sequence Number1
Product Code QLG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K193371
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date 08/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/25/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number71953-01
Was Device Available for Evaluation? No
Date Manufacturer Received08/14/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/26/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberQUALITY RECORD QR869779
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age48 YR
Patient SexFemale
-
-