• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST PAUL CADD EXTENSION SETS; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST PAUL CADD EXTENSION SETS; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number 21-7106-24
Device Problems Device Alarm System (1012); Obstruction of Flow (2423)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/28/2022
Event Type  malfunction  
Event Description
It was reported that a high pressure alarm was going off, and it appeared like no medical fluid had passed through the extension tube.No patient injury was reported.
 
Manufacturer Narrative
This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms#(b)(4).A product sample was received for evaluation.Visual inspection showed tamper seal was intact.Found high pressure, no disposable and power lost while pump was on messages in the event history log.During functional check, the reported complaint was duplicated.Cassette lock worn, case and lens damage issue found during investigation.Found front housing was damaged.Root cause was found to be the worn cassette lock, that was caused by the customer.The latch lock assembly, rear/front housing assembly and lens were replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD EXTENSION SETS
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key17012838
MDR Text Key316046127
Report Number3012307300-2023-05786
Device Sequence Number1
Product Code FPA
UDI-Device Identifier10610586023279
UDI-Public10610586023279
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K974013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 05/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number21-7106-24
Device Catalogue Number21-7106-24
Device Lot Number4220039
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/04/2022
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/06/2023
Initial Date FDA Received05/26/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/13/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-