U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TANDEM DIABETES CARE T:SLIM X2 INSULIN PUMP; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TANDEM DIABETES CARE T:SLIM X2 INSULIN PUMP; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number 1002717
Device Problems Incorrect Measurement (1383); Data Problem (3196)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/08/2023
Event Type  malfunction  
Manufacturer Narrative
The device is expected to be returned; however, the device has not yet been received.A supplemental report will be submitted if the device is received and evaluation is performed.H3 other text : device not returned.
 
Event Description
It was reported that the insulin gauge was inaccurate.During troubleshooting with tandem technical support (cts), it was determined that the customer was not removing the residual air from the cartridge correctly.Cts educated customer on proper load technique.The customer continued using the same cartridge.Customer¿s blood glucose level ranged from 108-145 mg/dl.In addition, it was reported that the pump indicated a data log corruption.Reportedly, customer continued using pump for insulin therapy.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
T:SLIM X2 INSULIN PUMP
Type of Device
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
TANDEM DIABETES CARE
12400 high bluff drive
san diego CA 92130
Manufacturer Contact
michael trier
san diego, CA 92130
8584011451
MDR Report Key17019720
MDR Text Key316119497
Report Number3013756811-2023-70503
Device Sequence Number1
Product Code OYC
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P140015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/30/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number1002717
Device Catalogue Number1002684
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Date Manufacturer Received07/17/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
INSULIN: NOVOLOG/NOVORAPID.
Patient Age22 YR
Patient SexFemale
Patient EthnicityNon Hispanic
-
-