TANDEM DIABETES CARE T:SLIM X2 INSULIN PUMP; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
|
Back to Search Results |
|
Model Number 1002717 |
Device Problems
Incorrect Measurement (1383); Data Problem (3196)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/08/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device is expected to be returned; however, the device has not yet been received.A supplemental report will be submitted if the device is received and evaluation is performed.H3 other text : device not returned.
|
|
Event Description
|
It was reported that the insulin gauge was inaccurate.During troubleshooting with tandem technical support (cts), it was determined that the customer was not removing the residual air from the cartridge correctly.Cts educated customer on proper load technique.The customer continued using the same cartridge.Customer¿s blood glucose level ranged from 108-145 mg/dl.In addition, it was reported that the pump indicated a data log corruption.Reportedly, customer continued using pump for insulin therapy.
|
|