BOSTON SCIENTIFIC CORPORATION RESONATE X4 CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D Back to Search Results
Model Number G447
Device Problems Signal Artifact/Noise (1036); Over-Sensing (1438); Pacing Problem (1439); Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problem Shock (2072)
Event Date 05/02/2023
Event Type  malfunction  
Manufacturer Narrative
Device remains implanted.If pertinent information is provided in the future, a supplemental report will be submitted.
 
Event Description
It was reported that this patient with a cardiac resynchronization therapy defibrillator (crt-d) device, received inappropriate anti-tachycardia pacing (atp) and 1 inappropriate shock, due to over-sensed noise on the rv lead.A technical service (ts) consultant reviewed device data, noting the noise is seen at the same time every night.The physician advised this is due to the patient using a vibrating plate on his leg at the same time as the events.Ts advised to discontinue using product due to emi interference.The device remains implanted.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESONATE X4 CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key17021550
MDR Text Key316141952
Report Number2124215-2023-27218
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P010012/S436
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/30/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/25/2019
Device Model NumberG447
Device Catalogue NumberG447
Device Lot Number164145
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/03/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/01/2017
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age88 YR
-
-