Model Number GF-UCT260
Device Problem Poor Quality Image (1408)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/05/2023
Event Type  malfunction  
Manufacturer Narrative
The device was returned, and an initial evaluation was conducted by olympus; however, investigation is ongoing.During the initial evaluation, the user¿s report was confirmed.There was part of the image missing because the adhesive of the probe was detached, and the probe was found blackened.Additionally, the unit had several other points of wear.The switches, channel tube, and protector were all found worn.The light guide bundle was damaged, the insertion tube was wrinkled, and the connector had oxidized.If additional information becomes available following the device evaluation, a supplemental report will be filed.
 
Event Description
The customer reported that while reprocessing his olympus evis lucera ultrasound gastrovideoscope, he discovered that the unit was producing an ultrasound image defect.There was no patient involvement during this event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the ultrasound image defect was due to user handling of the device.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS LUCERA ULTRASOUND GASTROVIDEOSCOPE
Type of Device
ULTRASOUND GASTROVIDEOSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17030829
MDR Text Key317222992
Report Number3002808148-2023-05436
Device Sequence Number1
Product Code ODG
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/31/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberGF-UCT260
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/09/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received06/01/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/04/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
CV-290
-
-