MEDOS INTERNATIONAL SARL FMS VUE II FLUID MANAGEMENT AND TISSUE DEBRIDEMENT SYSTEM; DISTENSION UNIT, FLUID, ARTHROSCOPIC
|
Back to Search Results |
|
Model Number 284004 |
Device Problems
Increase in Pressure (1491); Device Remains Activated (1525); Overfill (2404)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/12/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: udi: (b)(4).H4: the device manufacture date is currently unavailable.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
|
|
Event Description
|
It was reported by a healthcare professional that during an unknown procedure on (b)(6) 2023, a fms vue ii fluid management and tissue debridement system device was used.According to the report, the pump continued to run constantly and alarm.It was further reported that the device would run constantly that it would fill the chamber.It was further reported that its pressure was too high and too low.It was further reported that the solo constantly clicks unless door was open and then the orange light would come on.It was unknown if and how the procedure was completed.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
|
|
Manufacturer Narrative
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: the complaint device was received at the service center and evaluated.During the service evaluation, no defects were identified.Hence, this complaint cannot be confirmed.The device was found to be working according to the specifications.As part of depuy mitek¿s quality process all devices are manufactured, inspected, and released to approved specifications.Since the reported condition is not confirmed and no defects were identified, the root cause for the reported failure cannot be determined.The event described could not be confirmed as the device was returned without detectable damage.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the service evaluation.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Device history review: manufacturing record evaluation is not required as the reported event is not associated with the manufacturing process and/or the potential cause of the defect cannot be associated to manufacturing.
|
|
Search Alerts/Recalls
|
|
|