The patient was undergoing a coil embolization procedure in the varices using pod packing coils (pod pc), ruby coil, ruby coil detachment handle (handle), a lantern delivery microcatheter (lantern), a non-penumbra diagnostic catheter, and non-penumbra coils.During the procedure, the physician placed one ruby coil and two pod pcs in the target vessel using the lantern.The physician then advanced another pod pc into the vessel and detached it using the handle.However, while removing the pusher assembly of the pod pc, the physician noticed only part of the pusher assembly was removed.The physician proceeded to advance non-penumbra coils into the target location; however, the coils would not advance out from the lantern.The physician removed the lantern and flushed it on the back table but was not able to locate the non-penumbra coils.The physician then removed the diagnostic catheter and upon flushing it, the broken pusher assembly of the pod pc and other coils were removed from the catheter.The procedure was completed using a new pod pc and non-penumbra coils with the same handle, lantern, and diagnostic catheter.There was no adverse effect to the patient.
|
Evaluation of the returned pod pc confirmed that the pusher assembly was fractured and revealed kinks at the fractured locations.If the pusher assembly is manipulated against resistance during use, damage such as a kink and subsequent fracture may occur.Further evaluation revealed multiple kinks on the fractured segments.This damage was incidental to the reported complaint and the root cause could not be determined.Penumbra coils are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|