SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD ALARIS¿ PUMP MODULE SMARTSITE¿ INFUSION SET; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Model Number 2426-0007
Device Problems Leak/Splash (1354); Defective Component (2292)
Patient Problem Awareness during Anaesthesia (1707)
Event Date 04/19/2023
Event Type  malfunction  
Manufacturer Narrative
E.4.Fda notified?: the initial reporter also notified the fda via medwatch # mw5116969.H.3.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that the bd alaris¿ pump module smartsite¿ infusion set leaked propofol from the damaged tubing during use.This complaint was created to capture the 2nd of 3 related incidents.The following information was provided by the initial reporter: "multiple sets leaking - 1st report on 4/16, second report on 4/19, and 2 more verbal reports from nursing on 4/20 - good catches and interventions by nursing.The other 2 were verbals from nursing to house supervisor and then to me without details other than leaking tubing (not at the connection sites).Could you please clarify where exactly the leakage is taking place with the product in these events?: in the tubing itself.Were you able to identify if the product is leaking because of a damaged component or more of a loose connection?: damaged component.Was there any patient impact/harm in these events? if so was there any need for any medical intervention or treatment?: prior to finding leaking tubing different medications started for blood pressure dropping (once tubing replaced original medication restarted).Prior to finding leaking tubing on second one patient waking up from sedation meds not infusing.Able to correct.Was treatment able to resume and be completed in these events?: yes.Was any medication used with the product? if yes, what was the medication?: levophed was running with the first reported.Propofol was running with the second reported incident.Were there any concerns when priming the sets?: no".
 
Manufacturer Narrative
The following fields were updated due to additional information: d10: device available for evaluation? yes.D10: returned to manufacturer on: 28-jun-2023.H6: investigation summary 2 used sample (model#2426-0007) were returned for quality investigation by the customer.It was reported by the customer that the extensions tubing sets have been leaking.Prior to functional testing the sets were visually examined for defects and abnormalities.No other defects or abnormalities were observed.The set were connected to a 10 ml bd syringe filled with water and attempted to be flushed by pressure test.The fluid pass through the sample shows the leakage from the tubing.The leakage from tube was clearly observed because of hole in it was verified in magnification.Leakage issue could be replicated.The quality notification was sent to the plant.A device history record review for model 2426-0007 and lot number 23039059 was performed.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set of lot.After investigation, the potential root cause of damage tubing was not found to be related to the manufacturing process.
 
Event Description
It was reported that the bd alaris¿ pump module smartsite¿ infusion set leaked propofol from the damaged tubing during use.This complaint was created to capture the 2nd of 3 related incidents.The following information was provided by the initial reporter: "multiple sets leaking - 1st report on 4/16, second report on 4/19, and 2 more verbal reports from nursing on 4/20 - good catches and interventions by nursing.The other 2 were verbals from nursing to house supervisor and then to me without details other than leaking tubing (not at the connection sites).¿ could you please clarify where exactly the leakage is taking place with the product in these events? in the tubing itself.¿ were you able to identify if the product is leaking because of a damaged component or more of a loose connection? damaged component.¿ was there any patient impact/harm in these events? if so was there any need for any medical intervention or treatment? prior to finding leaking tubing different medications started for blood pressure dropping (once tubing replaced original medication restarted) prior to finding leaking tubing on second one patient waking up from sedation meds not infusing - - able to correct.¿ was treatment able to resume and be completed in these events? yes.¿ was any medication used with the product? if yes, what was the medication? levophed was running with the first reported - - - propofol was running with the second reported incident.¿ were there any concerns when priming the sets? no.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ALARIS¿ PUMP MODULE SMARTSITE¿ INFUSION SET
Type of Device
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8448235433
MDR Report Key17035301
MDR Text Key316305395
Report Number9616066-2023-01051
Device Sequence Number1
Product Code FPA
UDI-Device Identifier10885403227998
UDI-Public10885403227998
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K944320
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility
Reporter Occupation Nurse Practitioner
Type of Report Initial,Followup
Report Date 07/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/31/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2426-0007
Device Catalogue Number2426-0007
Device Lot Number23039059
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Manufacturer Received07/10/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/02/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-