AGILENT TECHNOLOGIES SINGAPORE (INTL.) PTE LTD. AUTOSTAINER LINK 48; AUTOMATED SLIDE STAINER Back to Search Results
Model Number AS480
Device Problems Fluid/Blood Leak (1250); Mechanical Problem (1384); Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/19/2023
Event Type  malfunction  
Manufacturer Narrative
This malfunction may potentially impact staining performance.No erroneous staining result was reported by the reporter with this incident.No patient or user harm was indicated.A1-a6: patient information has not been provided by the user.
 
Event Description
The spain customer reported "weak staining on some slides due to reagent not being dispensed properly".Fse (field service engineer) inspected the instrument and found aspiration and dispense check valve leaking.Fse replaced aspiration and dispense check valve which resolved this malfunction.The customer was able to complete staining.The instrument is fully operational, within specification, and ready for use.Diagnostics were not altered.There was no direct or indirect patient harm or user harm reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOSTAINER LINK 48
Type of Device
AUTOMATED SLIDE STAINER
Manufacturer (Section D)
AGILENT TECHNOLOGIES SINGAPORE (INTL.) PTE LTD.
no.1 yishun avenue 7
singapore north east, sgp 76892 3
SN  768923
Manufacturer (Section G)
SHANDON DIAGNOSTICS LIMITED
tudor road
manor park
runcorn, WA7 1 TA
UK   WA7 1TA
Manufacturer Contact
mary o'neill
1834 state highway 71 west
cedar creek, TX 78612
3026338510
MDR Report Key17037550
MDR Text Key316293515
Report Number3003423869-2023-00081
Device Sequence Number1
Product Code KPA
UDI-Device Identifier05700572035497
UDI-Public05700572035497
Combination Product (y/n)N
Reporter Country CodeSP
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Repair
Type of Report Initial
Report Date 06/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/01/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAS480
Device Catalogue NumberAS48030
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/22/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/07/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-