ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 1458Q/86
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/14/2023
Event Type  malfunction  
Event Description
It was reported during implant procedure on (b)(6) 2023 that the left ventricular lead exhibited lead damage.Further damage to the insulation was caused while attempting to advance the lead the guidewire.The lead was successfully exchanged.There were no patient consequences.
 
Manufacturer Narrative
The reported event of the lead insulation damage was confirmed.As received, a complete lead was returned in one piece for analysis.Visual inspection found the lead insulation was cut in the proximal region.The inner coil and ring electrode 1 cable were also damaged in this region.The cause of the reported event was consistent with procedural damage.
 
Manufacturer Narrative
Correction: updated b5 and h6 to reflect only insulation damage seen, no further damage noted.
 
Event Description
Additional information received notes that only insulation damage was noted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17050355
MDR Text Key316414256
Report Number2017865-2023-22121
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)Y
Reporter Country CodeIN
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 06/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/02/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1458Q/86
Device Catalogue Number1458Q-86
Device Lot NumberA000136698
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received06/27/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/20/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-