• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC EMAX 2 PLUS MOTOR; MOTOR, SURGICAL INSTRUMENT, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC EMAX 2 PLUS MOTOR; MOTOR, SURGICAL INSTRUMENT, AC-POWERED Back to Search Results
Model Number EMAX2PLUS
Device Problems Component Missing (2306); Fail-Safe Did Not Operate (4046)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2023
Event Type  malfunction  
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Additional narrative: reporter's phone number was not provided.The actual device has been returned and is currently pending evaluation.Once the evaluation of the device has been completed, a supplemental medwatch report will be sent accordingly.Udi (b)(4).
 
Event Description
It was reported from japan that the motor device was rotating in the locked position.It was reported that the connector pin was missing.It was not reported if the event occurred during a surgical procedure.It was not reported if there was a delay to a planned procedure.It was not reported if there was a spare device available for use.There was no patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was not reported, however, it was reported that the event occurred in 2023.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: the actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the motor device had cord damage/scratched, component damage, set screw was loose, electric leakage, damaged flex circuit, was making excessive noise, had vibration, the device would run in the locked position, and the etching was illegible.It was further determined that the device failed pretest for visual assessment, loctite assessment, no short circuit between phases and connector body, safety assessment, and noise assessment.Therefore, the reported condition was confirmed.The assignable root cause of this condition was determined to be traced to component failure due to wear.The reported condition that the connector pin was missing was not confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMAX 2 PLUS MOTOR
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
kate karberg
4500 riverside drive
palm beach gardens, FL 33410
3035526892
MDR Report Key17050514
MDR Text Key316420069
Report Number1045834-2023-00855
Device Sequence Number1
Product Code GEY
UDI-Device Identifier00845384002055
UDI-Public00845384002055
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K080802
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/02/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEMAX2PLUS
Device Catalogue NumberEMAX2PLUS
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/29/2023
Is the Reporter a Health Professional? No
Date Manufacturer Received06/05/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/17/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-