H4: device manufacture date: the lot was manufactured from september 1, 2022, to september 2, 2022.H10: the actual devices were not available; however, ten (10) companion samples were received for evaluation.A visual inspection with the unaided eye was performed on cavity b for all ten samples and it did not identify any abnormalities that could have contributed to the reported condition.System level testing was performed on two (2) randomly selected samples and the reported issue was not observed.Functional testing was performed with the two selected samples on by installing each valve set sample onto a compounder and a leak or air was not observed during the testing.Additionally, no leak was verified when attempting a ¿ui¿ flush after prime and during the pump calibration process.The reported condition was not verified on the companion samples.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|