Analysis of the returned device was completed on 09 jun 2023.159 therapies were conducted.The most recent therapy was on march 21, 2023.During two of these treatments, we monitored the temperature on the transducer surface, with the maximum temperature recorded reaching 25.3°c.Additionally, we measured acoustic power values both with and without the power supply connected, resulting in readings of 120 mw and 118 mw, respectively.No temperature-related issues were detected, and there was no confirmation of any malfunction with the device.
|