U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HILL-ROM/LIKO AB LIKO LIKORALL 243 ES; LIFT, PATIENT, NON-AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM/LIKO AB LIKO LIKORALL 243 ES; LIFT, PATIENT, NON-AC-POWERED Back to Search Results
Model Number 3123001NA
Device Problems Fail-Safe Problem (2936); Mechanics Altered (2984)
Patient Problem Fall (1848)
Event Date 05/25/2023
Event Type  Injury  
Event Description
On (b)(6) 2023.The ceiling lift failed during a transfer of the patient from wheelchair into bed.Safe patient handling coordinators went over to inspect the lift.Noted the lift would not go up, only down when the "up" button was pressed.Unit entered work order.On 5/26/23.(b)(6) spoke with the visn contract repair technician who reported the gear that bears weight is not working.If patient was lifted when the gear gave out, the patient would have dropped, like a previous incident.Tech removed lift from the room and staff advised to use portable patient lift.On 5/30/23.Liko rep notified and spoke on the phone.He agreed to do a complete sweep/inspection of the rest of the ceiling lifts (model 243, 507 bs.) to determine what the problem is.On 5/31/23.Liko rep came out to see the lift.He took a photo of the manufacturer sticker.We discussed a liko tech will be out to open up the lift for information.Awaiting date that tech will do the complete sweep/inspection of the rest of the ceiling lifts.Similarly, on (b)(6) 2021, with the same model of lift, nursing staff on the same spinal cord injury unit, applied sling to patient in a different room.The lift had worked just a few minutes before and they lowered it to attach the sling.With the patient all hooked up to the lift, they began to raise the lift and the lifting strap just gave way and dropped the patient on the bed.No time to reach for emergency cord.Patient was assessed for injury.No injury found.Safe patient handling coordinator spoke with staff, contract repair team, engineering, and liko vendor rep.No user error identified.Repair team identified the same issue with malfunctioning gear where the gears were just spinning instead of catching to stop the lift appropriately.These likorall 243 ceiling lifts were manufactured 2017 and installed february 2018.Reference report: mw5118159.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIKO LIKORALL 243 ES
Type of Device
LIFT, PATIENT, NON-AC-POWERED
Manufacturer (Section D)
HILL-ROM/LIKO AB
MDR Report Key17070481
MDR Text Key316698856
Report NumberMW5118158
Device Sequence Number1
Product Code FSA
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Nurse
Type of Report Initial
Report Date 06/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/05/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3123001NA
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-