Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d9: complainant part is not expected to be returned for manufacturer review/investigation.E3: reporter is a j&j employee.H3, h4, h6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
Device report from synthes reports an event in the united kingdom as follows: it was reported that on (b)(4) 2023, one of the transbuccal sets at the hospital had a threaded drill sleeve which may be slightly bent, since the transbuccal drill bits were not fitting properly.The sleeve and drill bit were heating up during drilling, indicating friction.A new instruments set was opened, and there was a delay of unknown length in the procedure.This report involves one matrixmandible threaded trocar drill guide for 397.213.This is report 1 of 2 for (b)(4).
|