Model Number N/A |
Device Problems
Component Missing (2306); Mechanical Jam (2983); Pressure Problem (3012)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/04/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
This complaint is recorded by zimmer biomet under (b)(4).A follow up/ final report will be submitted once investigation is complete if any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
It was reported that there was a turbine jam, no pressure and one screw is missing.There was no harm or delay reported.No adverse events were reported as a result of this malfunction.Due diligence is in progress.
|
|
Manufacturer Narrative
|
Following sections were updated or corrected :a3, b3, b5, d2, g1, g3, g6, h1, h2 and h6 if any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
Additional information was received and it was noted that the event timing was during surgery and there was a 16-30 minute delay.Due diligence is complete and no further information is available.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined the motor speed was unstable, a widthplate screw was missing, and the swivel was loose.The motor, sleeve, widthplate screws and swivel were replaced and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.G2: foreign: france.
|
|
Event Description
|
There is no additional information available.
|
|
Search Alerts/Recalls
|