• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOYA SURGICAL OPTICS, INC. B1PC; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOYA SURGICAL OPTICS, INC. B1PC; INTRAOCULAR LENS Back to Search Results
Model Number B1PC (+17.00 D)
Device Problem Device Handling Problem (3265)
Patient Problem Failure of Implant (1924)
Event Date 02/13/2023
Event Type  Injury  
Manufacturer Narrative
This initial emdr is being submitted to fda as a reportable event that occurred in the usa.Note: delayed emdr submission was due to fda esg webtrader account system software issues, per fda esg help desk ticket # (b)(4) (and subsequent ticket #(b)(4)).Hoya due diligence is documented under internal issue-0807.Injector malfunction is indicated as a potential malfunction related to the injector system, as covered under the warnings section of the product's instructions for use (ifu).Manufacturer's codes for: type of investigation, findings, and conclusion are pending device return and completion of product investigation.Once the product investigation is completed, a follow-up report will be submitted to fda which will include the manufacturer's codes for type of investigation, findings, and conclusion.
 
Event Description
Problem code: device handling problem.Injector malfunction was reported.Implant date: (b)(6) 2023; explant date: (b)(6) 2023 reported.
 
Event Description
Problem code: device handling problem.Injector malfunction was reported.Implant date: (b)(6) 2023; explant date: (b)(6) 2023 reported.
 
Manufacturer Narrative
This follow-up #1 emdr is being submitted to fda for a reportable event that occurred inside of the usa.The report includes additional information not available/included in the initial report.Additional information: h3 - indicated device not returned.G6 - type of report - noted as follow-up #1.H2 - type of follow-up - noted for additional information.H6 - added codes for manufacturer's investigation: type; findings; and conclusion.The product was not returned to the manufacturer.The investigation was conducted, with the methods and results as noted below.No abnormalities were found in production and inspection records of the product.(serial no.: (b)(6); model: b1pc) the exact root cause of the event was not determined.However, based on available information, we believe this event was not caused by our product quality.A review of the most recent complaint trending data indicates that no significant trends have been identified at this time and no capa is required as part of the product evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
B1PC
Type of Device
INTRAOCULAR LENS
Manufacturer (Section D)
HOYA SURGICAL OPTICS, INC.
110 progress
suite 175
irvine CA 92618
Manufacturer (Section G)
HOYA MEDICAL SINGAPORE PTE LTD
455a jalan ahmad ibrahim
singapore, singapore 63993 9
SN   639939
Manufacturer Contact
goutham pendyala
110 progress
suite 175
irvine, CA 92618
9093896317
MDR Report Key17080634
MDR Text Key316748962
Report Number3006723646-2023-00066
Device Sequence Number1
Product Code HQL
UDI-Device Identifier00886158017141
UDI-Public00886158017141
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P080004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative,Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/07/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2023
Device Model NumberB1PC (+17.00 D)
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received04/21/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured05/31/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-