It was reported, that during a procedure to treat atrial fibrillation.An intellanav mifi open-irrigated catheter was selected for use.The catheter could not reach the target location in the isthmus of the tricuspid ring.The bend of the ablation catheter was too large, resulting in a bend to the maximum angle, that could not reach the bottom of the isthmus.As such, the ablation effect was not satisfactory when it was applied.The catheter was replaced with another of the same model.The procedure was completed with no patient complications.However, analysis of the returned device revealed a cut in the catheter shaft.
|