ABBOTT QUARTET; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 1458Q/86
Device Problems Failure to Capture (1081); Fracture (1260); High impedance (1291)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/23/2023
Event Type  malfunction  
Event Description
It was reported that loss of capture and high pacing impedance were observed on the left ventricular (lv) lead.The suspected cause of the event was a lead fracture, however, this was not confirmed via diagnostic imaging.Abbott technical support was contacted and the device was reprogrammed to resolve the event.The patient was in stable condition and will continue to be monitored.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17110378
MDR Text Key317328093
Report Number2017865-2023-22829
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)Y
Reporter Country CodeIT
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 06/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/12/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2017
Device Model Number1458Q/86
Device Catalogue Number1458Q-86
Device Lot Number4703628
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/23/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/17/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-