• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 HEMOSTAR; DIALYSIS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 HEMOSTAR; DIALYSIS CATHETER Back to Search Results
Model Number 5833690
Device Problems Entrapment of Device (1212); Fracture (1260); Material Separation (1562); Stretched (1601); Unraveled Material (1664); Difficult to Advance (2920); Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/28/2023
Event Type  malfunction  
Manufacturer Narrative
H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation.However, a photo was provided for review.The investigation of the reported event is currently underway.H10: d4 (expiry date: 03/2024).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Event Description
It was reported that during a dialysis catheter placement procedure in the right atrium via the jugular vein, after the puncture, the guide wire allegedly met resistance while implanting and was difficult to adjust.It was further reported that when the guidewire was found difficult to adjust through the puncture needle, the whole device was removed to adjust, and it was found that the thin wire in the guide wire was allegedly fractured.There was no reported patient injury.
 
Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one 19cm hemostar d/l catheter, one 15.0fr peel-apart sheath and vessel dilator, one 8fr & 12fr dilator, one introducer needle, one j-tip guidewire, one guidewire hoop, one tunneler with a loaded protective sheath and one end cap were returned for evaluation.Gross visual, tactile, dimensional and functional evaluations were performed.The j-tip guidewire was noted to be uncoiled in the distal portion.Both core wires were protruding the coils of the guidewire and noted to have complete breaks on the ends.In addition to the returned physical sample, one electronic photo was provided for review.The photo shows a guidewire was stretched at the distal end and both the round and flat core wires were noted to be broken and protruding.Therefore, the investigation is confirmed for the reported fracture and identified material separation, stretched and unraveled issues.However, the investigation is inconclusive for the reported difficult to advance and physical resistance issues as the exact circumstances at the time of the reported event was unknown.Based on the measurements recorded during sample evaluation and applicable drawings, no measurements recorded could be confirmed to be out of tolerance.The definitive root cause could not be determined based upon the available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: d4 (expiration date: 03/2024).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that during a dialysis catheter placement procedure in the right atrium via the jugular vein, after the puncture, the guide wire allegedly met resistance while implanting and was difficult to adjust.It was further reported that when the guidewire was found difficult to adjust through the puncture needle, the whole device was removed to adjust, and it was found that the thin wire in the guide wire was allegedly fractured.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMOSTAR
Type of Device
DIALYSIS CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key17115176
MDR Text Key317041176
Report Number3006260740-2023-02400
Device Sequence Number1
Product Code MSD
UDI-Device Identifier00801741013386
UDI-Public(01)00801741013386
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K051748
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 02/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/13/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5833690
Device Catalogue Number5833690
Device Lot NumberREGT2532
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/25/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/05/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age65 YR
Patient SexMale
Patient Weight65 KG
-
-