The reported events of premature discharge of battery, no magnet response and no lv output were confirmed.The device was received with no telemetry and no output.Visual inspection of the header attachment area detected an anomaly between the pre-cast header and titanium case.The device was cut open to enable further testing and the battery was found below end-of-service voltage level.Feedthrough leak test was performed, indicating a device hermeticity breach.This is consistent with feedthrough damage as a result of fluid intrusion between the header and case, and subsequent fluid ingress to internal electronics.Hybrid circuitry was tested, and the results indicated elevated current drain, consistent with moisture damage, resulting in the reported events.A manufacturing process anomaly may have occurred, which resulted in the header bonding anomaly.
|