• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. HARMONIC ACE PLUS 7 ADV HEMOSTASIS 36; INSTRUMENT, ULTRASONIC SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. HARMONIC ACE PLUS 7 ADV HEMOSTASIS 36; INSTRUMENT, ULTRASONIC SURGICAL Back to Search Results
Model Number HARH36
Device Problem Tear, Rip or Hole in Device Packaging (2385)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2023
Event Type  malfunction  
Event Description
It was reported that during an unknown procedure the package was broken.Changed a new one and completed the surgery safely.There were no patient consequences.
 
Manufacturer Narrative
(b)(4).Date sent: 6/13/2023.B3: event year only reported: 2023.D4 batch #: x95348.Investigation summary the product was returned for evaluation.Visual inspection was conducted on the returned device.The device was returned sealed inside its original packaging.Upon visual inspection, it was observed that the blister from the packaging was damaged; it was noted to be broken and still adhered to the tyvek.Upon visual inspection, it was observed that the tyvek from the packaging was damaged.Due to the damages found on the blister, a possible cause for this condition is due to improper handling during transit or storage; it appears that the package hit a hard surface and this caused the reported event.Due to the damages found on the packaging, a possible cause for these conditions is due to improper handling during transit or storage; it appears that the package hit a pointy surface and this caused the reported event.The reported complaint was confirmed.All ees product is 100% inspected prior to release.The information you provided is compiled monitored and reviewed on a routine basis for any associated trends.As part of the quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device lot/batch x95348, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HARMONIC ACE PLUS 7 ADV HEMOSTASIS 36
Type of Device
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
*  
3035526892
MDR Report Key17118149
MDR Text Key317228393
Report Number3005075853-2023-04045
Device Sequence Number1
Product Code LFL
UDI-Device Identifier10705036014621
UDI-Public10705036014621
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K132612
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 06/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/13/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHARH36
Device Catalogue NumberHARH36
Device Lot NumberX95348
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/02/2023
Date Manufacturer Received06/05/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/31/2022
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
GENERATOR AND HANDPIECE
-
-