Model Number N/A |
Device Problem
Fracture (1260)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/22/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).H3- the product has been returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that a drill pin and screw inserter fractured.There is no additional information available.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Manufacturer Narrative
|
This follow-up is being submitted to relay additional information.Visual examination of the provided pictures identified the unit as broken.Additionally, the features of the fracture surface visually align with section 5.8.1 in zrm_wa_0545_21 in which an overload fracture failure mode was identified.Medical records were not provided.Device history record (dhr) was reviewed for deviations and/ or anomalies with no related deviations / anomalies identified.A definitive root cause cannot be determined.If any further information is received which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|