Catalog Number 8065750502 |
Device Problems
Device Emits Odor (1425); Operating System Becomes Nonfunctional (2996)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/16/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
|
Event Description
|
A customer reported that ophthalmic console emitted burning smell and turned off and was not turn on anymore.No patient harm was reported.
|
|
Manufacturer Narrative
|
The company representative was able to replicate the reported event.The power supply.Was replaced to address the issue.The system was tested and found to meet product specifications.A non-conformance based review of the serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for complaints reported against this serial number was performed.No similar complaints were reported for the product serial under investigation.The root cause of the reported event is attributed to nonconforming power supply.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
|
|
Search Alerts/Recalls
|