BIOTRONIK AG, BUELACH, SWITZERLAND PASSEO-14 2.5/140/150; PERIPHERAL DILATATION CATHETER
|
Back to Search Results |
|
Model Number 380339 |
Device Problem
Fluid/Blood Leak (1250)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/15/2023 |
Event Type
malfunction
|
Event Description
|
The passeo-14 balloon catheter was chosen for pta of a lower leg blood vessel.It was reported that the guidewire (no further information available) came out of the side port instead out of the guidewire lumen of the complaint instrument.Thus, a connection between the guidewire lumen and the inflation lumen was suspected.This was noted after using the balloon several times when the distal portion of the guidewire was still inside the patients body while the defective part of the instrument was outside of the patients body.During the preparations for the intervention, the complaint instrument could be normally flushed.The passeo-14 was used as usual in a standard way and no high pressure device was used.
|
|
Manufacturer Narrative
|
The returned product was subjected to a detailed technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.The technical investigation confirmed that the balloon has been inflated.The balloon was partially deflated in the as-returned state with a large amount of blood residue in the balloon- and inflation lumen.After properly flushing and cleaning of the instrument, a severe damage of the inner shaft inside the balloon became apparent.Microscopic inspection revealed that the inner shaft is sliced open over a length of about 9 mm, starting about 23 mm distal to the proximal x-ray marker.The inner shaft is both elongated and compressed distal and proximal to the sliced area.The distance between the two radiopaque markers does not comply anymore with the specification.A 0.014 inch reference guidewire could be normally introduced until reaching the sliced area.The guidewire used in the intervention was not returned for analysis.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.In addition to visual inspections each instrument is tested for air tightness by means of a pressure test and a helium leak test.We can therefore confirm that the instrument was delivered in a leak-proof condition.Based on the conducted investigations, no manufacturing or material related root cause could be determined.
|
|
Search Alerts/Recalls
|
|
|