Model Number 283512
Device Problems Break (1069); Overheating of Device (1437)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported by the affiliate in italy that preoperatively to an unknown procedure on an unknown date, it was observed that the fms tornado micro handpiece with buttons device had faulty parts.During in-house engineering evaluation, it was determined that the device had excessive heat.It was unknown if and how the procedure was completed.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: udi: (b)(4).E3: reporter is a j&j employee.Investigation summary: the complaint device was received at the service center and evaluated.During the service evaluation the following defects were identified: functional : excessive heat.Per service reports, this complaint can be confirmed.Since the repair quote was rejected by the customer, the device was not repaired.It was scrapped.As part of depuy mitek¿s quality process all devices are manufactured, inspected, and released to approved specifications.The faulty parts was identified as the root cause for the device failure during the service evaluation.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Device history review: manufacturing record evaluation is not required as the reported event is not associated with the manufacturing process and/or the potential cause of the defect cannot be associated to manufacturing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FMS TORNADO MICRO HANDPIECE WITH BUTTONS
Type of Device
LINE-POWERED ARTHROSCOPIC SHAVER SYSTEM
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key17131175
MDR Text Key317267027
Report Number1221934-2023-02419
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705016938
UDI-Public10886705016938
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K954465
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/14/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number283512
Device Catalogue Number283512
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/25/2023
Date Manufacturer Received05/31/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/19/2019
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-