• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ULTRATOME XL; SNARE, FLEXIBLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ULTRATOME XL; SNARE, FLEXIBLE Back to Search Results
Model Number M00535920
Device Problems Break (1069); Failure to Conduct (1114); Difficult to Remove (1528)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/09/2023
Event Type  malfunction  
Event Description
It was reported to boston scientific corporation that an ultratome xl was used in the ampulla and bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2023.During the procedure, the physician positioned the ultratome xl in the ampulla to perform sphincterotomy.He energized the wire; however, he noticed that the wire would not cut despite energizing the wire multiple times.The device was removed from the duodenoscope with difficulty, so he had to pull it hard.It was noticed that the cutting wire was broken.The elevator channel of the duodenoscope did not close anymore despite multiple efforts to close it and the duodenoscope had to be carefully extracted from the patient with the elevator channel in an open state.It was reported that no part of the cutting wire detached and fell into the patient.The procedure was not completed due to this event.There were no patient complications reported as a result of this event.The patient had to undergo a bile duct surgery, which was not related to this procedure.During that surgery, the stones present in the bile duct were removed.
 
Manufacturer Narrative
Block h6: imdrf device code a0401 captures the reportable event of cutting wire broken.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRATOME XL
Type of Device
SNARE, FLEXIBLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key17139610
MDR Text Key317767983
Report Number3005099803-2023-03222
Device Sequence Number1
Product Code FDI
UDI-Device Identifier08714729103318
UDI-Public08714729103318
Combination Product (y/n)N
Reporter Country CodeAU
PMA/PMN Number
K930022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 06/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/15/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM00535920
Device Catalogue Number3592
Device Lot Number0029629399
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/24/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/21/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-