Model Number 306546
Device Problem Pumping Problem (3016)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/22/2023
Event Type  malfunction  
Manufacturer Narrative
Investigation summary: it was reported the saline syringes alarmed "occluded" while running on syringe pump and were described as feeling "sticky" while flushing.To aid in the investigation, thirty-five samples in sealed packaging flow wraps were received for evaluation by our quality team.A visual inspection was performed, and no defects or imperfections were observed.Ten samples were randomly selected and tested for sustaining force, which is the force applied to the plunger rod while moving downwards when expelling the saline solution, and all results were within specification.It could be possible the customer is not using the product as intended.This product is not specified to work with a pump.If the product is not meeting the customer¿s needs, it may be beneficial to contact the local bd sales, or marketing representative, for additional product options.A device history record review was completed for provided material number 306546, lot 3068788.The review did not reveal any detected quality issues during the production of this lot that could have contributed to the reported defect.There were no related quality notifications.All processes and final inspections complied with specification requirements.To date, there have been no other similar events reported for this lot.Based on the investigation and with the returned sample analysis the symptom reported by the customer could not be confirmed.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.
 
Event Description
It was reported while using bd posiflush¿ normal saline syringes the pump alarms during use.There was no report of patient impact.The following information was provided by the initial reporter: 10ml normal saline syringes alarming "occluded" while running on syringe pump.Also described as feeling "sticky" while flushing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD POSIFLUSH¿ NORMAL SALINE SYRINGES
Type of Device
SALINE, VASCULAR ACCESS FLUSH
Manufacturer (Section D)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer (Section G)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8448235433
MDR Report Key17140549
MDR Text Key317945048
Report Number1911916-2023-00406
Device Sequence Number1
Product Code NGT
UDI-Device Identifier30382903065463
UDI-Public(01)30382903065463
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K003553
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,User Facility
Reporter Occupation Nurse
Type of Report Initial
Report Date 05/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/15/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number306546
Device Catalogue Number306546
Device Lot Number3068788
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/12/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/23/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/09/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-