ANGIOMED GMBH & CO. MEDIZINTECHNIK KG VENOVO VENOUS STENT Back to Search Results
Model Number VENUM20080
Device Problems Difficult or Delayed Positioning (1157); Activation Failure (3270)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/21/2023
Event Type  malfunction  
Manufacturer Narrative
H10: a voluntary recall has been initiated for the venovo¿ venous stent system 9f which was product catalog/lot number specific.Reportedly the proximal end of the stent does not immediately expand upon deployment.The proximal end of the stent remains connected to the delivery system.As a result of the field action, this event is being reported as a malfunction reportable event.H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the venovo venous stent products that are cleared in the us.The pro code and 510 k number for the venovo venous stent products are identified in d2 and g4.H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation.However, video was provided for review.The investigation of the reported event is currently underway.H10: d4 (expiry date: 09/2024).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.Device not returned.
 
Event Description
It was reported that during a stent placement procedure in superior vena cava via right brachial vein access, the proximal end of the stent allegedly failed to expand upon deployment.The procedure was completed using same device.There was no reported patient injury.
 
Manufacturer Narrative
H10: a voluntary recall has been initiated for the venovo¿ venous stent system 9f which was product catalog/lot number specific.Reportedly the proximal end of the stent does not immediately expand upon deployment.The proximal end of the stent remains connected to the delivery system.As a result of the field action, this event is being reported as a malfunction reportable event.H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the venovo venous stent products that are cleared in the us.The pro code and 510 k number for the venovo venous stent products are identified in d2 and g4.H10: manufacturing review: the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.Investigation summary: the sample was not returned for evaluation.One provided x-ray video demonstrates complete deployment of the stent inside the vessel; the proximal end of the stent is adhering to the inner catheter in the area of the stent cushion after complete stent sheath retraction.The video ends after approximately seven seconds of adhesion.High deployment force could not be visualized on the provided video.The investigation leads to confirmed result for stent expansion issue.Based on the information available the investigation is closed with confirmed result for stent expansion issue.A definite root cause for the reported event could not be determined.Placement in the vena cava is off label use.Labeling review: a review of the relevant instruction for use for this product was conducted.The instruction for use was found to address potential worst case complications and adverse events potentially related to expansion issue such as incorrect positioning of the stent requiring further stenting or surgery, intimal injury, stent fracture, and misplacement.The venovo venous stent system is indicated for the treatment of symptomatic iliofemoral venous outflow obstruction.H10: d4 (expiry date: 09/2024).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Event Description
It was reported that during a stent placement procedure in superior vena cava via right brachial vein access, the proximal end of the stent allegedly failed to expand upon deployment.The procedure was completed using same device.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENOVO VENOUS STENT
Type of Device
VENOUS STENT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM  76227
Manufacturer (Section G)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM   76227
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key17141895
MDR Text Key317295400
Report Number9681442-2023-00232
Device Sequence Number1
Product Code QAN
UDI-Device Identifier00801741103858
UDI-Public(01)00801741103858
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
P180037
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date 08/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/16/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberVENUM20080
Device Catalogue NumberVENEM20080
Device Lot NumberANGW3298
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/28/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-1897-2021
Patient Sequence Number1
-
-