Batch review performed on 25 may 2023 lot 2020841: (b)(4) items manufactured and released on 18-may-2020.Expiration date: 2025-may-03.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported event during the period of review.Other device involved: pedicle screw 03.56.504 must mc cross-connector 27-40mm adjustable (k193365) lot 2227010: (b)(4) items manufactured and released on 12-oct-2022.Expiration date: 2027-sep-22.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported event during the period of review.
|
The patient had a primary surgery on (b)(6) 2023, and had a revision surgery on (b)(6) 2023, due to cages migration.On (b)(6) 2023 surgery performed to extend the construct from l3-l5 to l3-s2ai.During the surgery, one screw in l5 was loose and the cross-connector in l5 was found with one hook loose.The cross-connector was removed and replaced with a new one, while the tulip of the screw was blocked, therefore the head adjuster had to be used to remove the screw because the screwdriver cannot be inserted in the head of the screw.The surgery was completed successfully.Delay time was 5min.
|
Visual inspection performed by r&d manager: ref.03.50.054 pedicle screw 8x40mm lot.2020841.Tulip is blocked with respect to the screw.The design of tulip for diameter >= 8mm features an opening, to allow assembly of the screw and give a bigger range of motion in one direction, useful for sai trajectory.This design may lead, during final fixation, to locking of the angle.No contraindications are foreseeable for the patient, if this happens.Loosening of the screw may be caused by several factors, such as bone quality, failed osteointegration, etc.Ref.03.56.504 must mc cross-connector 27-40mm adjustable lot.2227010 it was possible to firmly lock the connector in a test block.The potential root cause for the loosening may be the instability of the construct, given by the loosening of the screw.
|