• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TORNIER INC TORNIER HRS ASSEMBLY SCREW 0MM; PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TORNIER INC TORNIER HRS ASSEMBLY SCREW 0MM; PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED Back to Search Results
Catalog Number ARS655101
Device Problems Component Missing (2306); Device Markings/Labelling Problem (2911); Inaccurate Information (4051)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/24/2023
Event Type  malfunction  
Manufacturer Narrative
Analysis results are not available at the time of this report.A follow-up report will be sent when the analysis is complete.
 
Event Description
It was reported that the screw which was in the package was the wrong size.The size which should have been in place was a 0mm but this was even shorter and non-exsisting.
 
Event Description
It was reported that the screw which was in the package was the wrong size.The size which should have been in place was a 0mm but this was even shorter and non-exsisting.
 
Manufacturer Narrative
Correction: d9/h3, h6 component code.The reported could be confirmed.Photos were provided for investigation and clearly shows the discrepancy in the packaged device and what device should be in the package.A device inspection was not possible since the affected device was not returned, however photos were provided for investigation.The photos confirm that the screws provided for the case were too short for the proximal body to engage with the distal stem.There are two 0mm assembly screws marketed for the tornier hrs system.Catalog # ars655101 corresponds to the screw that is 36.5mm in length.Catalog # ars655118 corresponds to the ¿short¿ screw that is 26.5mm in length.The photos sent clearly show the screw is a ars655118 ¿0mm short screw¿ which will not function properly for this case.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.Based on the investigation, the root cause was attributed to a manufacturing related issue.It was discovered that ars655118 lot az1822082 (quantity (b)(4)) and ars655101 lot az1322077 (quantity (b)(4)) were processed through packaging at the same time.The supplier stated they believe this is a complete swap of lots.They have opened a capa on their end to formalize the root cause and escape.If any additional information is provided, the investigation will be reassessed.H3 other text : device not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TORNIER HRS ASSEMBLY SCREW 0MM
Type of Device
PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED
Manufacturer (Section D)
TORNIER INC
10801 nesbitt avenue s
bloomington MN 55437
Manufacturer (Section G)
TORNIER INC
10801 nesbitt avenue s
bloomington MN 55437
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key17146593
MDR Text Key317449867
Report Number0001649390-2023-00131
Device Sequence Number1
Product Code HSD
UDI-Device Identifier00846832093250
UDI-Public00846832093250
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
K181420
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/16/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberARS655101
Device Lot NumberAZ1322077
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received09/04/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/10/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-