• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL TACTICATH¿ CONTACT FORCE ABLATION CATHETER, SENSOR ENABLED¿ BID CURVE D-F; CARDIAC ABLATION PERCUTANEOUS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL TACTICATH¿ CONTACT FORCE ABLATION CATHETER, SENSOR ENABLED¿ BID CURVE D-F; CARDIAC ABLATION PERCUTANEOUS CATHETER Back to Search Results
Model Number A-TCSE-DF
Device Problems Display or Visual Feedback Problem (1184); No Apparent Adverse Event (3189)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
During a persistent atrial fibrillation procedure, after ablation, the signal on the distal end of the catheter would not show.The procedure was then stopped with no adverse patient consequences.
 
Event Description
During a persistent atrial fibrillation procedure, after ablation the signal on the distal end of the catheter would not show.The procedure was then stopped with no adverse patient consequences.The issue was isolated to the ensite x amplifier and there was no alleged issue with the tacticath se ablation catheter.
 
Manufacturer Narrative
Corrected information: b5, h6.Additional information: g3, h2, h3.During a persistent atrial fibrillation procedure, after ablation the signal on the distal end of the catheter would not show.The procedure was then stopped with no adverse patient consequences.The issue was isolated to the ensite x amplifier and there was no alleged issue with the tacticath se ablation catheter.The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TACTICATH¿ CONTACT FORCE ABLATION CATHETER, SENSOR ENABLED¿ BID CURVE D-F
Type of Device
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key17153065
MDR Text Key317874654
Report Number3008452825-2023-00262
Device Sequence Number1
Product Code OAE
UDI-Device Identifier05415067027641
UDI-Public05415067027641
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P130026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/19/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA-TCSE-DF
Device Catalogue NumberA-TCSE-DF
Device Lot Number8961447
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received08/02/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/26/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age69 YR
Patient SexFemale