• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. 2.5 MM FEMALE HEX SCREW 25 MM LENGTH; INSTRUMENT, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. 2.5 MM FEMALE HEX SCREW 25 MM LENGTH; INSTRUMENT, KNEE Back to Search Results
Model Number N/A
Device Problem Fracture (1260)
Patient Problem Foreign Body In Patient (2687)
Event Date 05/26/2023
Event Type  Injury  
Manufacturer Narrative
(b)(4).G2: foreign - japan.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch will be submitted.
 
Event Description
It was reported that during removal, the head of the hexagonal female was damaged and could not be removed.It remained in the patient.It was reported that no further information is available.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.The device history record was unable to be performed as the lot number of the device involved in the event is unknown.Review of complaint history found no additional related issues for this item and the reported part and lot combination.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.5 MM FEMALE HEX SCREW 25 MM LENGTH
Type of Device
INSTRUMENT, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key17154125
MDR Text Key317540845
Report Number0001822565-2023-01590
Device Sequence Number1
Product Code HWD
UDI-Device Identifier00889024232532
UDI-Public(01)00889024232532(17)330317(10)65766831
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/19/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number42509902525
Device Lot Number65766831
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/09/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/24/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Other;
Patient SexPrefer Not To Disclose
-
-