• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SWITZERLAND MANUFACTURER GMBH COCR HEAD 28/-4 S 12/14; HIP PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SWITZERLAND MANUFACTURER GMBH COCR HEAD 28/-4 S 12/14; HIP PROSTHESIS Back to Search Results
Catalog Number 14.28.05-20
Device Problems Device Dislodged or Dislocated (2923); Unintended Movement (3026); Migration (4003)
Patient Problem Joint Dislocation (2374)
Event Date 05/02/2023
Event Type  Injury  
Event Description
It was reported that patient underwent implantation of a hip prostheses.Postoperatively during exercise, the head and liner disassociated resulting in hip dislocation.Due diligence is in progress for this complaint; to date no additional information or product has been received.
 
Manufacturer Narrative
(b)(4).G2-foreign- germany.D10.Item#:61.27.28-42 ;lot#:unknown ;item name: bipolar insert, modular, 42/28; item#:4712500398-3 ;lot#:axdj2502 ;item name: optipac 40 refob bone cmt r-3 ; item#:01.06010.203 ;lot#: 3119158;item name: avenir, stem, standard, cemented, 3, taper 12/14 ; item#:61.27.00-44 ;lot#:2996323 ;item name: bipolar shell, modular, 44; multiple mdr reports were filed for this event, please see associated report: 0009613350 - 2023 - 00292 investigation of this incident is currently ongoing.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Event Description
It was reported that patient underwent implantation of a hip prostheses.Postoperatively during exercise, the head and liner disassociated resulting in hip dislocation, which required revision surgery.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.The following sections were updated: b2, b4, b5, d6, g3, g6, h2, h3, h6, h10.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.A review of the device manufacturing records for the cocr head confirmed no abnormalities or deviations that could be related to the reported event.A review of the device manufacturing records for the liner cannot be performed without product identification.Devices are used for treatment.The implant protocol for implantation surgery dated 2023 was received.Procedure performed was an implantation of a cemented bipolar prosthesis at the hip.The implanted products are provided.Of note is that the lot number of the liner is not provided.Based on the given information and results of the investigation, a definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COCR HEAD 28/-4 S 12/14
Type of Device
HIP PROSTHESIS
Manufacturer (Section D)
ZIMMER SWITZERLAND MANUFACTURER GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ  8404
Manufacturer (Section G)
ZIMMER SWITZERLAND MANUFACTURER GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ   8404
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key17161183
MDR Text Key317499028
Report Number0009613350-2023-00309
Device Sequence Number1
Product Code LPH
UDI-Device Identifier00889024393776
UDI-Public(01)00889024393776(17)321104(10)3138318
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/20/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number14.28.05-20
Device Lot Number3138318
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received08/04/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/27/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SEE H10 NARRATIVE
Patient Outcome(s) Required Intervention; Other; Hospitalization;
Patient Age79 YR
Patient SexFemale
-
-