• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX ANESTHESIA BREATHING CIRCUIT; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX ANESTHESIA BREATHING CIRCUIT; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE) Back to Search Results
Catalog Number CA2400/4A2/024JP
Device Problems Break (1069); Material Puncture/Hole (1504)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/01/2023
Event Type  malfunction  
Event Description
It was reported that before opening the package, the customer noticed the cuff was partially torn (or damaged).No adverse patient effects were reported by the customer.It was reported that no further information is available.
 
Manufacturer Narrative
B3: month and year of event have been provided, day is unknown.D4: lot number, expiration date, udi number and h4: device manufacture date is unknown, no information has been provided to date.G5: 510k is blank, device is exempt.One device was received in used pre-check condition without original packaging.Visual inspection noted the cuff was partially torn/damaged.The complaint was confirmed on visual inspection and no other analysis was performed.A root cause could not be determined however it is suspected the event occurred due to user interface after leaving the manufacturing facility.A device history record (dhr) review could not be performed as the lot number is unknown.While a root cause could not be determined awareness of the complaint has been made to production personnel.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX ANESTHESIA BREATHING CIRCUIT
Type of Device
CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key17162460
MDR Text Key317866367
Report Number3012307300-2023-06326
Device Sequence Number1
Product Code CAI
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 06/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/20/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCA2400/4A2/024JP
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/08/2023
Was the Report Sent to FDA? No
Date Manufacturer Received06/01/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-