Philips has investigated this complaint.According to the additional information collected, the system was in clinical use when the issue was identified.Due to the system issue, the planned treatment was delayed but no harm was reported.The philips remote service engineer (rse) inspected the system online and confirmed that system does not start.Review of the system log file showed that there was a malfunction with flexvision pc.Philips field service engineer (fse) visited onsite and noticed flexvision monitor had no layout lines on it.Then found that the flexvision pc didn¿t have any hard drive activity lights blinking.Fse replaced the flexvision pc, but the operating system installation failed.On a later date, fse visited again and completed the replacement of the flexvision pc and hard disk, reloaded the complete software.After replacement of flexvision pc, hard disk and installation of software, the system was returned to use in good working order.
|