• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION HOMECHOICE AUTOMATED PD SET WITH CASSETTE; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION HOMECHOICE AUTOMATED PD SET WITH CASSETTE; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number L5C4531
Device Problems Improper Flow or Infusion (2954); Protective Measures Problem (3015)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/28/2023
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a homechoice device experienced a system error 2240 (air in line/set) alarm.The patient was connected at the time of the alarm.This occurred during an unspecified cycle of peritoneal dialysis (pd) therapy.During the troubleshooting, it was reported that the ¿supply bag was swollen with air¿ during use of a homechoice cassette which led to the alarm.Renal therapy services advised the patient to contact their nurse regarding the issue.Proper procedures per the user manual were reviewed with the patient.The patient had manual supplies to continue therapy.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
H10: an alarm indicative of a potential malfunction of the disposable cassette was reported.The device was not returned; therefore, a device analysis could not be completed.However, it was reported that the ¿supply bag was swollen with air¿ during use of a homechoice cassette, which is known to cause this alarm.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - DOMINICAN REPUBLIC
carretera sanchez km 18.5
parque industrial itabo, piisa
haina, san cristobal 91000
DR   91000
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key17171796
MDR Text Key317578983
Report Number1416980-2023-03062
Device Sequence Number1
Product Code FKX
UDI-Device Identifier00085412090078
UDI-Public(01)00085412090078
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102936
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 07/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/21/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue NumberL5C4531
Device Lot NumberDR23B06062
Was Device Available for Evaluation? No
Date Manufacturer Received07/20/2023
Date Device Manufactured02/07/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DIANEAL PD-2 1.5%.; HOMECHOICE PRO DEVICE.
-
-