Additional information: d9, g3, h2, h3.One 5f, decapolar, medium sweep, livewire ep catheter was received for evaluation.Electrode ring 9 was dissected to reveal that conductor wire 9 was fractured proximal to the weld joint, consistent with the open circuit detected and the reported noise.Electrodes 1-2 and 10 met specifications of acceptable resistance values with no open or short circuits detected.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the fractured conductor wire and subsequent delay remains unknown.
|