• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION CYSTOSCOPY IRRIGATION SET; SYSTEM, IRRIGATION, UROLOGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION CYSTOSCOPY IRRIGATION SET; SYSTEM, IRRIGATION, UROLOGICAL Back to Search Results
Catalog Number 2C4040
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/11/2023
Event Type  malfunction  
Event Description
It reported that there was hair in the tubing of a cysto/bladder irrigation set.The issue was identified during set-up.There was no patient involvement.
 
Manufacturer Narrative
G1: device manufacturer address 1: 600 mts.Oeste de entrada, principal ave.Las americas, parque industrial.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
H10: the device was received for evaluation.A visual inspection was done which observed particulate matter (pm) on the tubing (not inside the tube).The particulate has the following characteristics: black and solid state, and according to laboratory analysis it was identified as a hair.The reported condition was verified.The cause of the condition was due to a manufacturing issue related to human production.Since the pm was found to be outside the fluid path, this issue is not likely to cause or contribute to a death or serious injury.Pm outside of the fluid path does not impact the product delivered to the patient.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CYSTOSCOPY IRRIGATION SET
Type of Device
SYSTEM, IRRIGATION, UROLOGICAL
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - CARTAGO
see h10
see h10
cartago 30106
CS   30106
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key17173191
MDR Text Key318199206
Report Number1416980-2023-03078
Device Sequence Number1
Product Code LJH
UDI-Device Identifier00085412001050
UDI-Public(01)00085412001050
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K960787
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 07/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/21/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2C4040
Device Lot NumberR22F22147
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Manufacturer Received07/26/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/23/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-