BAXTER HEALTHCARE CORPORATION CYSTOSCOPY IRRIGATION SET; SYSTEM, IRRIGATION, UROLOGICAL
|
Back to Search Results |
|
Catalog Number 2C4040 |
Device Problem
Device Contamination with Chemical or Other Material (2944)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/11/2023 |
Event Type
malfunction
|
Event Description
|
It reported that there was hair in the tubing of a cysto/bladder irrigation set.The issue was identified during set-up.There was no patient involvement.
|
|
Manufacturer Narrative
|
G1: device manufacturer address 1: 600 mts.Oeste de entrada, principal ave.Las americas, parque industrial.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
H10: the device was received for evaluation.A visual inspection was done which observed particulate matter (pm) on the tubing (not inside the tube).The particulate has the following characteristics: black and solid state, and according to laboratory analysis it was identified as a hair.The reported condition was verified.The cause of the condition was due to a manufacturing issue related to human production.Since the pm was found to be outside the fluid path, this issue is not likely to cause or contribute to a death or serious injury.Pm outside of the fluid path does not impact the product delivered to the patient.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|