It was reported that, when priming the tubing with medication, the tubing would not prime past the filter, the pump got to 10ml and automatically stopped.It was supposed to prime by 7mll.No patient injury reported.
|
Other, other text: one sample was received for evaluation.Visual inspection found no obstructions, damage, breakage, or other defects detected.To conduct functional testing a priming test was performed.No discrepancies were detected, sample was fully priming and connected without difficulty, the pump was set running, liquid flow through the whole device without interruptions, no alarms were activated, thus, the failure mode reported is not confirmed.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.
|