MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. CADD HIGH VOLUME ADMINISTRATION SET WITH FLOWSTOP; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. CADD HIGH VOLUME ADMINISTRATION SET WITH FLOWSTOP; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number 21-7346-24
Device Problem Priming Problem (4040)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/29/2023
Event Type  malfunction  
Manufacturer Narrative
A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported that, when priming the tubing with medication, the tubing would not prime past the filter, the pump got to 10ml and automatically stopped.It was supposed to prime by 7mll.No patient injury reported.
 
Manufacturer Narrative
Other, other text: one sample was received for evaluation.Visual inspection found no obstructions, damage, breakage, or other defects detected.To conduct functional testing a priming test was performed.No discrepancies were detected, sample was fully priming and connected without difficulty, the pump was set running, liquid flow through the whole device without interruptions, no alarms were activated, thus, the failure mode reported is not confirmed.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD HIGH VOLUME ADMINISTRATION SET WITH FLOWSTOP
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key17174073
MDR Text Key317634064
Report Number3012307300-2023-06480
Device Sequence Number1
Product Code FPA
UDI-Device Identifier15019517161953
UDI-Public15019517161953
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K031361
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/21/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number21-7346-24
Device Catalogue Number21-7346-24
Device Lot Number4320132
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/30/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/19/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-