• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TORAX MEDICAL, INC. UNK LINX MAGNETIC IMPLANT; ANTI-REFLUX IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TORAX MEDICAL, INC. UNK LINX MAGNETIC IMPLANT; ANTI-REFLUX IMPLANT Back to Search Results
Catalog Number UNK LINX MAGNETIC IMPLANT
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Patient Device Interaction Problem (4001)
Patient Problems Pyrosis/Heartburn (1883); Nausea (1970); Pain (1994)
Event Date 06/14/2023
Event Type  Injury  
Event Description
It was reported that a patient who had linx device placed 2 years ago with concomitant hiatal hernia repair via cruroplasty has had excellent response with only complaint being globus sensation.However, in the past several months experiencing reflux, nausea, postprandial pain.Upper gi series shows retention of contrast in esophagus secondary to significant narrowing at the linx device.Upper gi endoscopy was performed showing the narrowing and balloon dilatation was performed.The patient did not have any response to the dilatation and continues with the same symptoms without any improvement.No additional information was provided.
 
Manufacturer Narrative
(b)(4).Date sent: 6/22/2023.B3: unknown; captured as awareness date.D6a: exact implant date is unk.Implant year is 2021.Assumed first month of the year and first day of the month.No lot number was provided therefore a device history could not be done.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNK LINX MAGNETIC IMPLANT
Type of Device
ANTI-REFLUX IMPLANT
Manufacturer (Section D)
TORAX MEDICAL, INC.
4545 creek road
cincinnati OH
Manufacturer Contact
kate karberg
4545 creek rd.
cincinnati, OH 45242
3035526892
MDR Report Key17180814
MDR Text Key317633718
Report Number3008766073-2023-00112
Device Sequence Number1
Product Code LEI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100049
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional
Reporter Occupation Other
Type of Report Initial
Report Date 06/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNK LINX MAGNETIC IMPLANT
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 06/14/2023
Initial Date FDA Received06/22/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age28 YR
Patient SexFemale
-
-