Ultra ice plus catheter was evaluated by boston scientific.Visual inspection noted kinks were observed in the imaging window assembly.An image and impedance inspection were performed.Impedance testing shows a good unit wave form.During image characterization testing, a good square image appeared in the ilab system.-microscope inspection did not note any damage consistent with saline exposure post-use of the device.Finally, a functional inspection was done and it was observed that the catheter leaks from the imaging window section.Although laboratory analysis was unable to confirm the reported clinical observations, device presented other failure findings.
|