Device evaluation summary: the reported liquid qc results issue was verified during service.Technician was unable to duplicate the reported problem, however observed 56 error 10 (bad ref), lift wire height was too high at.1175 (spec=.0880-.0960).This could cause the error 10 and issues with failing controls.The issue was resolved by setting the lift wire height to.0895/.0940.The tilt frame was replaced as a precautionary measure for parallelism.Ran electronic and liquid controls with no problems noted.Preventive maintenance was performed per specification.Conclusion: the complaint is confirmed for the act plus instruments liquid qc results issue.No patient/clinical safety issues reported.Trends for issues with this product are reviewed at quarterly quality meetings.The act plus software constantly monitors for software and/or hardware faults.When one is found, an error is displayed and typically addressed by the operator.No further actions to be taken at this time.This regulatory report is being submitted as part of a retrospective review and remediation per (b)(4) as part of a capa action.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|