TERUMO CARDIOVASCULAR SYSTEMS CORPORATION STERILE FX15RWC W/ 3L RES; BLOOD GAS OXYGENATOR Back to Search Results
Model Number 3CX*FX15RW30C
Device Problem Unintended Movement (3026)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/01/2023
Event Type  malfunction  
Manufacturer Narrative
Terumo has received the device for evaluation.However, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.For this reason, terumo references evaluation conclusion code 11.H6: component code: 918 - probe.Health effect - impact code: 2199 - no health consequences or impact.Health effect - clinical code: 4582 - no clinical signs, symptoms or conditions.Medical device problem code: 3026 - unintended movement.Investigation findings: 3233 - results pending completion of investigation.Investigation conclusions: 11 - conclusion not yet available.
 
Event Description
The user facility reported to terumo cardiovascular that during cardiopulmonary bypass, the venous inlet thermistor was loosen.No known impact or consequence to patient.The product was not changed out.The surgery was completed successfully.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on june 23, 2023.Upon further investigation of the reported event, the following information is new and/or changed: d4 (additional device information - added expiration date); g3 (date received by manufacturer); g6 (indication that this is a follow-up report); h2 (follow-up due to additional information); h3 (device evaluated by manufacturer) ; h4 (device manufacture date); h6 (identification of evaluation codes 10, 11, 3331, 3259, 4307).Type of investigation #1: 10 - testing of actual/suspected device.Type of investigation #2: 11 - testing of device from same lot/batch retained by manufacturer.Type of investigation #3: 3331 - analysis of production records.Investigation finding: 3259 - improper physical structure.Investigation conclusions: 4307 - cause traced to component failure.The returned sample was inspected and confirmed that the venous temp probe was broken off the port.It was also noted that there was significant damage to the end of the venous inlet port.A representative retention sample was inspected for damage with no damage noted on the device.All capiox units are 100% visually inspected at several points in the production process.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Event Description
Loose venous inlet thermistor.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERILE FX15RWC W/ 3L RES
Type of Device
BLOOD GAS OXYGENATOR
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
Manufacturer (Section G)
SAME
Manufacturer Contact
douglas patton
125 blue ball road
elkton, MD 21921
7346634145
MDR Report Key17190248
MDR Text Key317772386
Report Number1124841-2023-00157
Device Sequence Number1
Product Code DTZ
UDI-Device Identifier00699753450783
UDI-Public(01)00699753450783
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K151389
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 07/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/23/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3CX*FX15RW30C
Device Catalogue NumberN/A
Device Lot Number2L12
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/20/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received07/21/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/16/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-