• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number V60
Device Problems Volume Accuracy Problem (1675); Failure to Deliver (2338)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/12/2023
Event Type  malfunction  
Manufacturer Narrative
Reporter phone number - (b)(6).
 
Event Description
Philips received a complaint by the customer on the v60 indicating that the tidal volume output was 0.The device was in use on a patient at the time the reported issue was discovered.However, there was no harm to the patient or user.The customer evaluated the device with the assistance of the remote service engineer (rse) and confirmed the reported problem.The investigation is ongoing.
 
Manufacturer Narrative
H10: the customer called in to report that an alarm occurred indicating that the tidal volume output was 0.The customer's equipment department then cleaned the flow sensor and confirmed the issue was resolved.The customer's equipment department cleaned the flow sensor to resolve the reported issue.The device passed required performance verification tests per philips standards and was returned to service.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key17190591
MDR Text Key317745117
Report Number2518422-2023-14066
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838025776
UDI-Public00884838025776
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/23/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Device Catalogue Number1076716
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/25/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-